Devises for measuring parametres of units and ciruits with lumped constants (immittance meters)
IMMITTANCE METER E7-15
The meter is used for measuring immittance parameters of electric and radio components at laboratories, workshops and repair shops.
Application:
- incoming and production control of the parameters of capacitors, resistors, inductance coils, transformers, relays, etc.
- immittance investigation of various structural components: transformers, screens, selector switches;
- measurement of different physical quantities such as humidity, pressure, solution strength, linear dimensions, by means transducers.
Design features: extremely simple to operate, automatic selection of measurement ranges, self-diagnostics when the meter is connected to the mains (check on ROM contents, on main memory serviceability, on general functioning of the ratio-meter); when the instrument is under repair, its assemblies may be checked more comprehensively in the testing mode to localize the trouble to a faulty stage (section). The built-in single-chip computer controls all the instrument assemblies, makes all calculations. Four-pair connection of the components under test into the measuring circuit provides high metrological characteristics.
The measurement result is presented on the display.
| SPECIFICATIONS: |
| Operating frequencies, kHz |
0,1; 1 |
| Measurement range: |
|
| Capacitance (C), F |
0,1x10-12... 20x10-3 |
| Inductance (L) Hen |
0,1x10-6... 1,6x103 |
| Resistance (R) Ohm |
1x10-3...20x106 |
| Conductance (G) Sm |
1x10-9...0,1x10-6 |
| Loss factor |
1x10-3...2 |
| Quality factor |
1x10-3 ... 0,5 |
| Measurement error, % |
±0,25 |
| Measuring signal Level, V |
2±0,4 (root-mean-square value) |
| AC power supply, V/Hz |
220/50; 220/400 |
| Power consumption, VA |
20 |
| Dimensions, mm |
254x184x314 |
| Weight, kg, max |
5 |
| Category temperature range, °C |
- 10 up to 40 |
|
|
|
^
Copyright © 2003 РУП "Минский завод Калибр"
|